*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Jan 2008 14:58:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Jan 2008 14:59:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 080110 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 236.35 stddev = 5.56 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_080110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1723.8 1723.8 1359.7 0.13 0.14 1.27 2 8.51 3415.0 3414.7 2874.5 -0.12 -0.11 1.19 3 17.02 6797.3 6797.3 5227.2 -0.22 -0.22 1.30 4 25.53 10186.2 10185.5 7750.4 -0.20 -0.20 1.31 5 38.29 15283.1 15282.8 11481.7 -0.09 -0.09 1.33 6 51.06 20385.1 20384.7 15095.8 -0.01 -0.01 1.35 7 63.83 25496.3 25496.9 18599.4 0.07 0.07 1.37 8 85.10 34014.8 34015.6 24068.9 0.16 0.16 1.41 9 106.38 42518.9 42520.3 29215.2 0.17 0.17 1.46 10 127.66 50987.4 50987.1 34256.3 0.11 0.11 1.49 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2574 to 76125 e- Average count rate measured: 399 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.49 +/- 0.02 e-/ADU RON = 8.30 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.066 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay