*** MIDAS logfile (05SEPpl1.2) *** We, 12 Dec 2007 14:05:26 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 12 Dec 2007 14:05:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 071212 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 238.62 stddev = 4.48 FIERA temperature = 283.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_071212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1751.2 1751.5 1373.7 0.14 0.14 1.27 2 8.51 3478.8 3478.8 2692.3 -0.13 -0.14 1.29 3 17.02 6935.6 6936.0 5526.4 -0.22 -0.22 1.25 4 25.53 10398.3 10398.3 7878.5 -0.20 -0.20 1.32 5 38.29 15608.2 15608.3 11750.4 -0.08 -0.08 1.33 6 51.06 20821.8 20822.1 15307.8 -0.01 -0.00 1.36 7 63.83 26045.7 26046.7 19295.3 0.08 0.08 1.35 8 85.10 34749.7 34749.7 24609.0 0.16 0.16 1.41 9 106.38 43432.9 43432.7 29865.0 0.16 0.16 1.45 10 127.66 52087.1 52087.5 34700.1 0.10 0.10 1.50 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2622 to 77981 e- Average count rate measured: 407 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.03 e-/ADU RON = 6.70 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.040 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay