*** MIDAS logfile (05SEPpl1.2) *** Su, 14 Oct 2007 15:45:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 Oct 2007 15:45:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 071014 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 295.08 stddev = 4.59 FIERA temperature = 285.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_071014 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1810.9 1810.9 1445.3 0.05 0.05 1.25 2 8.51 3582.6 3582.9 2930.8 -0.06 -0.05 1.22 3 17.02 7125.8 7125.0 5608.8 -0.10 -0.11 1.27 4 25.53 10671.8 10672.1 8248.5 -0.09 -0.09 1.29 5 38.29 16002.9 16002.4 12107.1 -0.01 -0.01 1.32 6 51.06 21334.2 21333.7 15959.2 0.04 0.03 1.34 7 63.83 26670.1 26671.7 19588.6 0.08 0.09 1.36 8 85.10 35556.0 35555.5 25479.5 0.10 0.10 1.40 9 106.38 44411.2 44411.3 30911.5 0.05 0.04 1.44 10 127.66 53226.8 53227.7 35933.5 -0.07 -0.07 1.48 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2680 to 78787 e- Average count rate measured: 417 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.48 +/- 0.02 e-/ADU RON = 6.79 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.090 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay