*** MIDAS logfile (05SEPpl1.2) *** Mo, 05 Mar 2007 16:46:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 05 Mar 2007 16:46:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 070305 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 302.62 stddev = 4.77 FIERA temperature = 285.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_070305 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1883.8 1883.9 1497.8 0.04 0.05 1.26 2 8.51 3728.3 3728.0 2918.1 -0.04 -0.05 1.28 3 17.02 7415.6 7415.3 5756.8 -0.09 -0.10 1.29 4 25.53 11107.6 11107.5 8554.7 -0.07 -0.07 1.30 5 38.29 16655.2 16655.3 12619.7 -0.00 -0.00 1.32 6 51.06 22205.2 22204.9 16574.7 0.05 0.05 1.34 7 63.83 27759.2 27759.5 20433.0 0.09 0.09 1.36 8 85.10 37004.3 37005.5 26483.7 0.10 0.10 1.40 9 106.38 46213.4 46213.8 31961.0 0.03 0.03 1.45 10 127.66 55381.9 55379.0 36813.0 -0.09 -0.10 1.50 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2815 to 82757 e- Average count rate measured: 434 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.49 +/- 0.03 e-/ADU RON = 7.13 +/- 0.15 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.088 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay