*** MIDAS logfile (05SEPpl1.2) *** We, 03 Jan 2007 16:06:41 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 03 Jan 2007 16:06:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 070103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 303.17 stddev = 4.80 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_070103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1897.7 1898.0 1519.1 0.05 0.05 1.25 2 8.51 3773.3 3773.3 2968.4 -0.04 -0.04 1.27 3 17.02 7522.1 7522.1 5869.7 -0.09 -0.09 1.28 4 25.53 11275.8 11275.9 8799.4 -0.07 -0.07 1.28 5 38.30 16916.6 16916.8 12795.7 0.01 0.01 1.32 6 51.06 22558.2 22558.3 16770.1 0.05 0.05 1.35 7 63.83 28204.8 28203.8 20661.3 0.10 0.09 1.37 8 85.10 37602.0 37602.0 26708.0 0.10 0.10 1.41 9 106.38 46955.1 46957.9 32275.3 0.01 0.02 1.45 10 127.66 56266.9 56269.1 37303.7 -0.12 -0.12 1.51 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2854 to 84622 e- Average count rate measured: 441 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.03 e-/ADU RON = 7.22 +/- 0.14 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.049 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay