*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 18:05:11 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 18:06:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 226.39 stddev = 4.80 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1906.9 1906.7 1474.4 0.17 0.16 1.29 2 8.51 3792.0 3792.2 2954.6 -0.15 -0.15 1.28 3 17.02 7564.8 7563.6 5740.9 -0.27 -0.28 1.32 4 25.53 11347.6 11347.6 8544.8 -0.22 -0.22 1.33 5 38.29 17035.8 17037.6 12694.3 -0.10 -0.09 1.34 6 51.06 22736.0 22736.0 16661.6 0.01 0.01 1.36 7 63.83 28443.5 28443.1 20444.1 0.10 0.10 1.39 8 85.10 37957.1 37957.0 26459.4 0.19 0.19 1.43 9 106.38 47434.0 47435.3 31915.8 0.17 0.18 1.49 10 127.66 56878.9 56879.0 36765.8 0.10 0.10 1.55 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2933 to 87485 e- Average count rate measured: 445 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.54 +/- 0.03 e-/ADU RON = 7.39 +/- 0.16 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.026 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.27 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay