*** MIDAS logfile (05SEPpl1.2) *** Sa, 01 Jul 2006 15:34:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 01 Jul 2006 15:35:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 060701 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 225.48 stddev = 4.68 FIERA temperature = 288.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_060701 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1933.6 1933.4 1502.4 0.16 0.17 1.29 2 8.51 3839.7 3839.7 2944.2 -0.17 -0.16 1.30 3 17.02 7655.4 7655.4 5779.7 -0.27 -0.27 1.32 4 25.53 11481.4 11481.4 8601.7 -0.22 -0.22 1.33 5 38.30 17236.8 17236.4 12714.8 -0.09 -0.10 1.36 6 51.06 23000.2 23000.7 17002.3 0.01 0.01 1.35 7 63.83 28775.6 28774.3 20615.7 0.11 0.10 1.40 8 85.10 38394.9 38395.5 26684.8 0.20 0.19 1.44 9 106.38 47982.1 47981.7 32427.9 0.18 0.17 1.48 10 127.66 57529.3 57528.3 37127.7 0.10 0.09 1.55 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2970 to 88380 e- Average count rate measured: 450 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.54 +/- 0.03 e-/ADU RON = 7.19 +/- 0.16 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.035 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.27 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay