*** MIDAS logfile (05SEPpl1.2) *** Mo, 20 Feb 2006 17:37:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 20 Feb 2006 17:39:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 060220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 293.17 stddev = 4.86 FIERA temperature = 289.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_060220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 2012.7 2012.5 1595.7 0.05 0.05 1.26 2 8.51 3991.4 3991.3 3145.4 -0.04 -0.04 1.27 3 17.02 7945.7 7945.7 6131.7 -0.11 -0.11 1.30 4 25.53 11906.8 11906.0 9072.2 -0.08 -0.08 1.31 5 38.30 17859.4 17859.4 13588.2 0.01 0.01 1.31 6 51.06 23814.2 23814.9 17597.0 0.06 0.06 1.35 7 63.83 29774.3 29774.8 21675.3 0.11 0.11 1.37 8 85.10 39690.3 39690.9 27957.6 0.11 0.11 1.42 9 106.38 49559.3 49558.4 33475.8 0.02 0.02 1.48 10 127.66 59376.7 59377.3 38623.5 -0.13 -0.13 1.54 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3081 to 90899 e- Average count rate measured: 465 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.53 +/- 0.03 e-/ADU RON = 7.45 +/- 0.17 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.069 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay