*** MIDAS logfile (05SEPpl1.2) *** Sa, 21 Jan 2006 15:33:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 21 Jan 2006 15:35:13 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 060121 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 321.47 stddev = 5.60 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_060121 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 1038.8 1038.9 862.5 0.10 0.10 1.20 2 8.00 2062.3 2062.0 1637.3 -0.09 -0.10 1.26 3 16.00 4111.0 4111.0 3357.5 -0.14 -0.14 1.22 4 24.00 6161.6 6161.5 4814.7 -0.13 -0.13 1.28 5 36.00 9240.4 9240.3 7198.8 -0.09 -0.09 1.28 6 48.00 12322.9 12323.2 9433.2 -0.04 -0.03 1.31 7 60.00 15408.9 15408.0 11751.5 0.02 0.01 1.31 8 80.00 20554.2 20554.2 15352.4 0.08 0.08 1.34 9 100.00 25704.3 25705.0 18917.3 0.13 0.14 1.36 10 120.00 30851.9 30851.4 22229.8 0.16 0.17 1.39 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1442 to 42809 e- Average count rate measured: 257 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.39 +/- 0.01 e-/ADU RON = 7.76 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.047 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay