*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 16:14:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 16:14:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 080209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 276.91 stddev = 5.37 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1662.7 1662.1 1346.6 -0.13 -0.10 1.23 2 2.12 3231.1 3231.3 2596.7 0.20 0.23 1.24 3 4.25 6340.2 6340.0 5020.9 -0.02 -0.01 1.26 4 6.38 9462.6 9462.2 7547.1 0.04 0.04 1.25 5 9.57 14140.1 14139.4 11149.4 0.05 0.04 1.27 6 12.76 18828.3 18828.3 14833.0 0.11 0.09 1.27 7 15.95 23500.3 23502.4 18731.6 0.08 0.07 1.25 8 21.27 31281.0 31282.8 24128.8 0.02 0.01 1.30 9 26.59 39033.0 39033.0 31475.4 -0.09 -0.10 1.24 10 31.91 46737.1 46736.2 35246.5 -0.26 -0.27 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2157 to 60633 e- Average count rate measured: 1465 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.02 e-/ADU RON = 6.97 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.076 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.27 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay