*** MIDAS logfile (05SEPpl1.2) *** We, 30 Jan 2008 17:02:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 30 Jan 2008 17:03:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 080130 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 274.97 stddev = 5.46 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_080130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1754.7 1754.0 1425.0 -0.11 -0.13 1.23 2 2.12 3376.6 3377.3 2716.0 0.24 0.24 1.24 3 4.25 6592.5 6589.1 5230.2 0.01 -0.01 1.26 4 6.38 9820.2 9820.7 7879.5 0.06 0.06 1.25 5 9.57 14654.5 14654.0 11869.1 0.05 0.05 1.23 6 12.76 19503.1 19502.2 15423.2 0.11 0.11 1.26 7 15.95 24334.7 24333.6 19026.8 0.08 0.08 1.28 8 21.27 32373.4 32372.4 24866.4 -0.00 0.00 1.30 9 26.59 40382.3 40384.4 31977.4 -0.12 -0.11 1.26 10 31.91 48336.5 48340.3 36232.6 -0.31 -0.30 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2306 to 63550 e- Average count rate measured: 1515 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.02 e-/ADU RON = 7.18 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.100 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.30 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay