*** MIDAS logfile (05SEPpl1.2) *** Th, 03 Jan 2008 15:39:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 03 Jan 2008 15:40:10 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 080103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 277.26 stddev = 5.60 FIERA temperature = 288.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_080103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1668.0 1668.9 1355.7 -0.14 -0.14 1.23 2 2.12 3244.0 3244.6 2615.1 0.21 0.24 1.24 3 4.25 6368.3 6367.9 5066.6 -0.02 -0.02 1.26 4 6.38 9505.9 9505.4 7687.4 0.05 0.03 1.24 5 9.57 14207.3 14206.3 11309.7 0.05 0.05 1.26 6 12.76 18916.9 18920.0 14882.7 0.10 0.11 1.27 7 15.95 23614.2 23615.3 18561.5 0.08 0.08 1.27 8 21.27 31435.0 31433.3 24320.3 0.03 0.02 1.29 9 26.59 39218.8 39222.9 29868.1 -0.10 -0.09 1.31 10 31.91 46961.5 46959.1 35371.7 -0.26 -0.28 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2215 to 62335 e- Average count rate measured: 1472 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 7.43 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.076 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.23/-0.27 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay