*** MIDAS logfile (05SEPpl1.2) *** We, 12 Dec 2007 14:32:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 12 Dec 2007 14:32:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 071212 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.92 stddev = 4.44 FIERA temperature = 283.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_071212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1676.0 1676.5 1354.1 -0.13 -0.12 1.24 2 2.12 3261.5 3262.6 2616.7 0.22 0.23 1.25 3 4.25 6405.9 6404.5 5090.6 -0.01 -0.02 1.26 4 6.38 9563.4 9561.2 7628.4 0.06 0.04 1.25 5 9.57 14289.4 14290.5 11279.6 0.05 0.05 1.27 6 12.76 19030.2 19031.7 14987.4 0.10 0.10 1.27 7 15.95 23755.6 23753.9 18780.6 0.08 0.07 1.26 8 21.27 31619.8 31619.7 24759.8 0.01 0.01 1.28 9 26.59 39454.2 39456.2 31439.4 -0.10 -0.10 1.25 10 31.91 47236.7 47240.4 35895.0 -0.28 -0.28 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2170 to 61153 e- Average count rate measured: 1481 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 5.75 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.074 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.23/-0.28 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay