*** MIDAS logfile (05SEPpl1.2) *** Su, 14 Oct 2007 16:11:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 Oct 2007 16:11:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 071014 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 315.97 stddev = 5.08 FIERA temperature = 285.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_071014 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1832.3 1831.9 1532.3 -0.21 -0.19 1.20 2 2.12 3518.6 3517.6 2848.3 0.26 0.27 1.24 3 4.25 6856.1 6854.9 6212.4 0.04 0.04 1.10 4 6.38 10208.8 10207.4 8148.6 0.10 0.10 1.25 5 9.57 15228.5 15226.5 12038.9 0.10 0.09 1.26 6 12.76 20257.5 20257.6 16138.4 0.14 0.13 1.26 7 15.95 25267.4 25269.1 20199.8 0.09 0.09 1.25 8 21.27 33604.7 33604.6 26342.9 -0.01 -0.02 1.28 9 26.59 41905.9 41907.2 32496.1 -0.15 -0.15 1.29 10 31.91 50145.2 50144.0 38618.8 -0.36 -0.37 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2390 to 65424 e- Average count rate measured: 1572 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 6.63 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.108 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.37 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay