*** MIDAS logfile (05SEPpl1.2) *** Th, 27 Sep 2007 18:23:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 27 Sep 2007 18:23:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 070927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 317.66 stddev = 5.31 FIERA temperature = 282.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_070927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1718.3 1719.4 1405.9 -0.17 -0.14 1.22 2 2.12 3335.7 3335.8 2688.9 0.28 0.28 1.24 3 4.25 6541.0 6541.8 5257.3 0.03 0.04 1.24 4 6.38 9760.4 9759.2 7800.8 0.09 0.07 1.25 5 9.57 14579.5 14580.0 11768.8 0.07 0.07 1.24 6 12.76 19413.0 19412.4 15461.3 0.12 0.11 1.26 7 15.95 24225.3 24223.0 19160.6 0.08 0.07 1.26 8 21.27 32233.1 32234.5 25363.0 -0.02 -0.02 1.27 9 26.59 40215.7 40215.4 31637.6 -0.14 -0.14 1.27 10 31.91 48132.3 48138.8 36641.2 -0.35 -0.34 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2235 to 62593 e- Average count rate measured: 1510 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 6.90 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.080 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.34 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay