*** MIDAS logfile (05SEPpl1.2) *** Sa, 28 Apr 2007 18:12:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 28 Apr 2007 18:12:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 070428 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 321.12 stddev = 6.67 FIERA temperature = 282.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_070428 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1732.1 1732.5 1439.9 -0.20 -0.20 1.20 2 2.12 3383.8 3383.2 2780.7 0.26 0.26 1.22 3 4.25 6654.1 6652.8 5360.6 0.04 0.03 1.24 4 6.38 9939.5 9937.8 7989.8 0.10 0.09 1.24 5 9.57 14856.0 14856.4 11869.8 0.08 0.08 1.25 6 12.76 19786.8 19785.4 15690.2 0.13 0.13 1.26 7 15.95 24697.5 24696.9 20305.0 0.09 0.09 1.22 8 21.27 32867.4 32868.2 25586.3 -0.01 -0.00 1.28 9 26.59 41006.0 41006.3 33079.2 -0.14 -0.13 1.24 10 31.91 49083.5 49083.8 37310.0 -0.35 -0.34 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2236 to 63343 e- Average count rate measured: 1540 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.02 e-/ADU RON = 8.61 +/- 0.14 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.067 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.35 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay