*** MIDAS logfile (05SEPpl1.2) *** Th, 29 Mar 2007 15:16:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 29 Mar 2007 15:17:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 070329 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 320.63 stddev = 5.25 FIERA temperature = 284.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_070329 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1732.5 1733.5 1457.5 -0.20 -0.17 1.19 2 2.12 3391.5 3391.5 2757.2 0.29 0.28 1.23 3 4.25 6674.6 6674.7 5359.3 0.03 0.04 1.25 4 6.38 9972.1 9971.3 8051.5 0.09 0.09 1.24 5 9.57 14909.1 14910.7 12028.6 0.08 0.08 1.24 6 12.76 19859.0 19859.0 15842.5 0.13 0.13 1.25 7 15.95 24788.1 24788.8 19695.5 0.08 0.08 1.26 8 21.27 32989.4 32990.0 25472.6 -0.01 -0.02 1.30 9 26.59 41157.7 41159.8 31571.7 -0.15 -0.15 1.30 10 31.91 49269.4 49267.6 40680.0 -0.35 -0.36 1.21 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2175 to 61822 e- Average count rate measured: 1546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.02 e-/ADU RON = 6.59 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.063 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.36 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay