*** MIDAS logfile (05SEPpl1.2) *** We, 14 Mar 2007 18:27:39 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 14 Mar 2007 18:27:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 070314 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 324.88 stddev = 5.26 FIERA temperature = 282.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_070314 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1887.4 1887.2 1572.2 -0.22 -0.18 1.20 2 2.12 3640.0 3639.5 2953.4 0.29 0.30 1.23 3 4.25 7108.6 7109.4 5746.6 0.04 0.06 1.24 4 6.38 10591.5 10592.9 8459.2 0.10 0.11 1.25 5 9.57 15809.9 15808.6 12859.2 0.10 0.09 1.23 6 12.76 21040.1 21038.8 16746.7 0.15 0.14 1.26 7 15.95 26247.0 26246.3 20678.7 0.09 0.08 1.27 8 21.27 34907.7 34908.6 27177.8 -0.01 -0.02 1.28 9 26.59 43528.0 43530.6 34324.3 -0.17 -0.18 1.27 10 31.91 52084.8 52086.4 39483.6 -0.39 -0.40 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2465 to 68039 e- Average count rate measured: 1634 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.01 e-/ADU RON = 6.87 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.097 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.30/-0.40 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay