*** MIDAS logfile (05SEPpl1.2) *** Mo, 05 Mar 2007 17:20:37 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 05 Mar 2007 17:20:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 070305 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.84 stddev = 5.82 FIERA temperature = 285.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_070305 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1911.5 1911.3 1556.7 -0.17 -0.16 1.23 2 2.12 3671.2 3669.8 3275.6 0.30 0.29 1.12 3 4.25 7150.9 7151.3 5750.3 0.04 0.05 1.24 4 6.38 10650.8 10649.6 8522.4 0.12 0.11 1.25 5 9.57 15886.2 15886.9 12778.8 0.09 0.10 1.24 6 12.76 21135.7 21135.3 16774.6 0.14 0.14 1.26 7 15.95 26360.1 26362.0 20663.5 0.08 0.09 1.28 8 21.27 35056.3 35056.2 27326.4 -0.02 -0.02 1.28 9 26.59 43707.1 43709.3 33387.0 -0.18 -0.17 1.31 10 31.91 52295.7 52296.6 39186.5 -0.41 -0.41 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2548 to 69701 e- Average count rate measured: 1640 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 7.76 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.107 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.29/-0.41 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay