*** MIDAS logfile (05SEPpl1.2) *** We, 03 Jan 2007 16:34:48 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 03 Jan 2007 16:35:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 070103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.90 stddev = 5.24 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_070103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1773.1 1770.7 1472.2 -0.22 -0.22 1.20 2 2.12 3467.6 3465.8 3294.5 0.30 0.30 1.05 3 4.25 6815.9 6814.4 5482.8 0.04 0.04 1.24 4 6.38 10180.4 10181.5 8628.5 0.10 0.11 1.18 5 9.57 15219.5 15218.3 12774.8 0.09 0.08 1.19 6 12.76 20269.8 20272.8 16213.4 0.13 0.14 1.25 7 15.95 25300.5 25301.6 19822.8 0.09 0.09 1.28 8 21.27 33671.5 33672.7 26144.5 -0.01 -0.02 1.29 9 26.59 42003.2 42005.9 32238.5 -0.15 -0.15 1.30 10 31.91 50277.5 50283.2 38300.4 -0.36 -0.36 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2351 to 66726 e- Average count rate measured: 1578 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 6.95 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.065 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.30/-0.36 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay