*** MIDAS logfile (05SEPpl1.2) *** Th, 14 Dec 2006 17:15:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 14 Dec 2006 17:16:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 061214 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 320.73 stddev = 5.13 FIERA temperature = 287.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_061214 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1919.6 1920.8 1573.5 -0.17 -0.22 1.22 2 2.12 3698.3 3698.8 2993.2 0.29 0.28 1.24 3 4.25 7215.3 7215.8 5807.0 0.05 0.05 1.24 4 6.38 10750.4 10750.6 8547.4 0.11 0.12 1.26 5 9.57 16041.1 16040.9 13368.9 0.10 0.10 1.20 6 12.76 21344.6 21344.7 16765.8 0.14 0.15 1.27 7 15.95 26623.8 26625.4 20780.1 0.09 0.10 1.28 8 21.27 35408.9 35407.5 28964.1 -0.02 -0.01 1.22 9 26.59 44150.0 44149.8 34361.6 -0.18 -0.17 1.28 10 31.91 52825.2 52824.5 39474.4 -0.41 -0.40 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2520 to 69334 e- Average count rate measured: 1657 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.02 e-/ADU RON = 6.73 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.101 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.40 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay