*** MIDAS logfile (05SEPpl1.2) *** Th, 07 Dec 2006 15:30:03 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 07 Dec 2006 15:30:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 061207 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.67 stddev = 5.43 FIERA temperature = 288.4 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_061207 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1917.4 1917.4 1581.2 -0.23 -0.16 1.21 2 2.12 3699.4 3697.5 2962.5 0.30 0.29 1.25 3 4.25 7220.1 7220.0 5785.2 0.05 0.06 1.25 4 6.38 10758.9 10757.8 8589.4 0.11 0.10 1.25 5 9.57 16055.6 16054.4 12802.9 0.10 0.09 1.25 6 12.76 21366.4 21366.0 17197.4 0.15 0.14 1.24 7 15.95 26653.8 26654.4 20924.6 0.10 0.09 1.27 8 21.27 35449.4 35448.6 28294.5 -0.01 -0.02 1.25 9 26.59 44202.0 44201.9 33688.8 -0.16 -0.18 1.31 10 31.91 52886.2 52885.6 39441.8 -0.40 -0.41 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2545 to 70199 e- Average count rate measured: 1659 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.02 e-/ADU RON = 7.21 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.098 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.29/-0.41 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay