*** MIDAS logfile (05SEPpl1.2) *** We, 15 Nov 2006 15:21:14 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 15 Nov 2006 15:21:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 061115 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.59 stddev = 5.15 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_061115 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1938.3 1939.6 1582.1 -0.21 -0.19 1.23 2 2.13 3731.2 3730.8 3049.4 0.27 0.27 1.22 3 4.25 7278.9 7277.7 6128.2 0.07 0.06 1.19 4 6.38 10839.6 10838.4 8679.3 0.12 0.11 1.25 5 9.57 16172.4 16172.5 12882.3 0.10 0.10 1.26 6 12.76 21516.8 21517.9 16866.0 0.14 0.15 1.28 7 15.96 26841.2 26841.4 21043.8 0.10 0.10 1.28 8 21.28 35693.3 35695.9 27929.6 -0.01 -0.01 1.28 9 26.59 44505.3 44506.8 34144.6 -0.17 -0.17 1.30 10 31.91 53242.3 53243.4 40364.1 -0.41 -0.41 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2558 to 70256 e- Average count rate measured: 1670 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 6.79 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.102 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.41 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay