*** MIDAS logfile (05SEPpl1.2) *** Su, 29 Oct 2006 16:38:36 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 29 Oct 2006 16:39:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 061029 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.51 stddev = 4.82 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_061029 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1926.8 1927.3 1565.7 -0.19 -0.19 1.23 2 2.12 3714.7 3714.3 3051.6 0.27 0.25 1.22 3 4.25 7254.0 7253.9 6194.8 0.05 0.05 1.17 4 6.38 10809.0 10809.4 8712.2 0.12 0.12 1.24 5 9.57 16129.7 16129.6 12761.4 0.10 0.10 1.26 6 12.76 21465.5 21466.3 17273.1 0.15 0.15 1.24 7 15.95 26777.4 26778.0 21070.3 0.10 0.10 1.27 8 21.27 35612.5 35612.1 27463.5 -0.01 -0.01 1.30 9 26.59 44406.7 44408.0 33858.0 -0.17 -0.17 1.31 10 31.91 53130.3 53137.0 39968.9 -0.40 -0.39 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2569 to 70824 e- Average count rate measured: 1666 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 6.42 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.099 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.39 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay