*** MIDAS logfile (05SEPpl1.2) *** We, 18 Oct 2006 15:17:49 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 18 Oct 2006 15:18:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 061018 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.78 stddev = 4.81 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_061018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1788.2 1788.1 1463.4 -0.20 -0.15 1.22 2 2.12 3496.9 3495.6 2856.7 0.28 0.28 1.22 3 4.25 6876.1 6876.9 5596.7 0.03 0.04 1.23 4 6.38 10271.9 10271.5 8143.1 0.09 0.09 1.26 5 9.57 15357.8 15358.1 12222.4 0.09 0.08 1.26 6 12.76 20454.5 20456.2 16180.0 0.14 0.13 1.26 7 15.95 25532.8 25533.3 20047.6 0.09 0.08 1.27 8 21.27 33978.0 33981.1 26994.1 -0.01 -0.02 1.26 9 26.59 42392.1 42389.9 32608.3 -0.14 -0.16 1.30 10 31.91 50735.6 50737.8 38130.9 -0.37 -0.38 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2358 to 66912 e- Average count rate measured: 1593 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.02 e-/ADU RON = 6.34 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.064 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.37 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay