*** MIDAS logfile (05SEPpl1.2) *** Fr, 06 Oct 2006 14:40:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 06 Oct 2006 14:41:17 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 061006 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.97 stddev = 4.54 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_061006 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1937.9 1937.5 1656.1 -0.22 -0.16 1.17 2 2.12 3735.9 3735.6 3006.4 0.26 0.28 1.24 3 4.25 7297.0 7294.8 5884.1 0.08 0.06 1.24 4 6.38 10870.2 10870.1 8653.1 0.12 0.12 1.26 5 9.57 16220.7 16221.0 12857.3 0.10 0.10 1.26 6 12.76 21586.1 21587.7 16979.6 0.15 0.15 1.27 7 15.95 26927.7 26928.5 21233.4 0.10 0.09 1.27 8 21.27 35811.7 35811.6 28954.0 -0.01 -0.02 1.24 9 26.59 44652.2 44651.4 34012.3 -0.17 -0.19 1.31 10 31.91 53419.4 53420.2 40049.9 -0.41 -0.43 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2559 to 70557 e- Average count rate measured: 1676 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.02 e-/ADU RON = 6.00 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.098 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.42 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay