*** MIDAS logfile (05SEPpl1.2) *** We, 06 Sep 2006 15:53:17 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Sep 2006 15:54:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060906 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 282.02 stddev = 4.98 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060906 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1785.2 1785.4 1495.1 -0.15 -0.10 1.19 2 2.12 3490.2 3489.1 2796.2 0.22 0.23 1.25 3 4.25 6866.5 6867.3 5541.4 -0.03 -0.02 1.24 4 6.38 10261.8 10260.5 9073.2 0.05 0.04 1.13 5 9.57 15344.1 15344.6 12112.4 0.05 0.05 1.27 6 12.76 20442.7 20441.7 16059.1 0.13 0.11 1.27 7 15.95 25521.4 25522.0 19853.2 0.09 0.08 1.29 8 21.27 33971.9 33973.5 26239.0 0.02 0.01 1.29 9 26.59 42392.9 42393.2 31758.9 -0.09 -0.10 1.33 10 31.91 50750.8 50752.5 37830.5 -0.29 -0.30 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2412 to 68574 e- Average count rate measured: 1592 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.02 e-/ADU RON = 6.73 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.062 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.30 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay