*** MIDAS logfile (05SEPpl1.2) *** Th, 24 Aug 2006 20:38:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 24 Aug 2006 20:40:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060824 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 282.42 stddev = 5.58 FIERA temperature = 288.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060824 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1924.1 1923.0 1562.8 -0.19 -0.13 1.23 2 2.12 3708.1 3707.2 2970.5 0.26 0.28 1.25 3 4.25 7236.4 7235.5 5745.0 -0.00 -0.00 1.26 4 6.38 10783.8 10783.2 8476.1 0.07 0.06 1.27 5 9.57 16094.0 16093.5 12613.0 0.07 0.06 1.28 6 12.76 21420.2 21421.7 16742.0 0.13 0.12 1.28 7 15.95 26726.6 26728.3 20715.2 0.10 0.09 1.29 8 21.27 35554.1 35554.4 27446.9 0.02 0.00 1.30 9 26.59 44341.0 44342.0 33291.7 -0.12 -0.13 1.33 10 31.91 53064.4 53062.2 39364.3 -0.33 -0.35 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2587 to 71356 e- Average count rate measured: 1663 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 7.51 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.098 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.34 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay