*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Aug 2006 14:25:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Aug 2006 14:26:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060810 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 280.12 stddev = 4.94 FIERA temperature = 287.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060810 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1795.5 1795.5 1449.7 -0.14 -0.12 1.24 2 2.12 3508.0 3507.5 2808.8 0.24 0.23 1.25 3 4.25 6899.0 6897.6 5467.9 -0.03 -0.03 1.26 4 6.38 10308.6 10306.6 8083.8 0.06 0.04 1.28 5 9.57 15412.4 15412.1 12100.0 0.06 0.06 1.27 6 12.76 20534.6 20530.4 15945.3 0.13 0.12 1.29 7 15.95 25628.7 25627.9 20079.5 0.08 0.08 1.28 8 21.27 34118.0 34116.0 26205.4 0.02 0.01 1.30 9 26.59 42566.5 42571.3 32079.3 -0.11 -0.10 1.33 10 31.91 50960.0 50964.0 38056.5 -0.31 -0.30 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2402 to 68178 e- Average count rate measured: 1599 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 6.60 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.065 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.30 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay