*** MIDAS logfile (05SEPpl1.2) *** Sa, 05 Aug 2006 18:38:05 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 05 Aug 2006 18:39:01 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060805 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 280.85 stddev = 5.13 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060805 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1941.9 1940.8 1559.9 -0.16 -0.14 1.24 2 2.12 3732.3 3731.9 3014.7 0.23 0.25 1.24 3 4.25 7277.6 7277.6 6590.1 -0.00 0.00 1.10 4 6.38 10841.2 10840.2 8685.4 0.06 0.06 1.25 5 9.57 16176.4 16176.4 13279.9 0.07 0.06 1.22 6 12.76 21528.6 21527.8 16729.2 0.14 0.13 1.29 7 15.95 26857.8 26855.8 20971.5 0.10 0.09 1.28 8 21.27 35723.4 35725.1 27413.6 0.01 0.01 1.30 9 26.59 44552.6 44549.4 34259.4 -0.12 -0.13 1.30 10 31.91 53309.4 53312.2 39316.1 -0.34 -0.35 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2621 to 71973 e- Average count rate measured: 1671 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.02 e-/ADU RON = 6.92 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.103 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.34 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay