*** MIDAS logfile (05SEPpl1.2) *** Tu, 11 Jul 2006 15:57:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Tu, 11 Jul 2006 15:58:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060711 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 281.03 stddev = 4.71 FIERA temperature = 287.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060711 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1938.8 1934.7 1567.8 -0.15 -0.19 1.24 2 2.12 3744.4 3743.4 3009.8 0.25 0.24 1.24 3 4.25 7314.9 7316.6 5799.5 0.00 0.01 1.26 4 6.38 10903.1 10904.5 8637.5 0.07 0.08 1.26 5 9.57 16276.0 16277.1 12871.3 0.06 0.07 1.26 6 12.76 21667.5 21666.9 16985.0 0.14 0.14 1.28 7 15.95 27034.5 27034.7 22013.1 0.10 0.10 1.23 8 21.27 35961.8 35963.5 27464.8 0.01 0.01 1.31 9 26.59 44850.5 44855.0 33795.1 -0.13 -0.12 1.33 10 31.91 53673.3 53671.9 39867.1 -0.35 -0.35 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2600 to 72046 e- Average count rate measured: 1683 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.02 e-/ADU RON = 6.33 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.095 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.35 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay