*** MIDAS logfile (05SEPpl1.2) *** Th, 06 Jul 2006 13:10:56 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 06 Jul 2006 13:12:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060706 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 280.50 stddev = 5.07 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060706 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1837.9 1839.0 1477.1 -0.17 -0.14 1.24 2 2.12 3586.5 3585.4 2860.2 0.27 0.26 1.25 3 4.25 7042.6 7042.7 5566.1 -0.01 -0.01 1.27 4 6.38 10516.5 10516.6 8297.8 0.06 0.05 1.27 5 9.57 15720.2 15720.1 12360.8 0.06 0.06 1.27 6 12.76 20938.6 20937.6 16304.1 0.12 0.12 1.28 7 15.95 26135.9 26137.7 20255.6 0.09 0.09 1.29 8 21.27 34788.1 34786.6 26729.7 0.01 0.01 1.30 9 26.59 43398.6 43398.2 32957.7 -0.12 -0.12 1.32 10 31.91 51951.7 51951.2 38789.7 -0.32 -0.32 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2454 to 69351 e- Average count rate measured: 1630 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 6.77 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.070 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.32 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay