*** MIDAS logfile (05SEPpl1.2) *** Su, 28 May 2006 15:35:48 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 28 May 2006 15:36:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060528 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 324.69 stddev = 5.10 FIERA temperature = 288.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060528 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1979.4 1978.5 1634.5 -0.20 -0.16 1.21 2 2.13 3815.6 3816.1 3092.5 0.24 0.27 1.23 3 4.25 7454.8 7454.2 6024.7 0.07 0.06 1.24 4 6.38 11109.9 11110.4 8880.9 0.13 0.13 1.25 5 9.57 16578.8 16579.2 13264.0 0.11 0.10 1.25 6 12.76 22063.7 22063.3 17317.5 0.16 0.15 1.27 7 15.95 27523.4 27525.4 21652.2 0.11 0.10 1.27 8 21.28 36601.6 36601.8 28325.2 -0.01 -0.02 1.29 9 26.59 45631.5 45635.9 34861.2 -0.18 -0.19 1.31 10 31.91 54589.7 54593.5 40591.7 -0.43 -0.44 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2646 to 72981 e- Average count rate measured: 1713 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 6.82 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.096 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.43 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay