*** MIDAS logfile (05SEPpl1.2) *** Fr, 19 May 2006 21:06:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 19 May 2006 21:07:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060519 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 326.24 stddev = 5.35 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060519 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1840.2 1839.5 1518.4 -0.17 -0.19 1.21 2 2.13 3602.7 3602.7 2954.8 0.27 0.27 1.22 3 4.25 7093.5 7093.3 5682.1 0.05 0.05 1.25 4 6.38 10601.2 10601.8 9674.0 0.11 0.12 1.10 5 9.57 15849.6 15847.7 12559.9 0.09 0.09 1.26 6 12.76 21111.5 21113.2 16793.9 0.14 0.14 1.26 7 15.95 26351.7 26352.6 20740.4 0.09 0.09 1.27 8 21.27 35069.5 35068.3 27285.1 -0.02 -0.02 1.29 9 26.59 43745.7 43744.5 33240.2 -0.17 -0.17 1.32 10 31.91 52355.0 52355.9 39708.7 -0.40 -0.40 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2448 to 69668 e- Average count rate measured: 1644 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.02 e-/ADU RON = 7.12 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.060 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.40 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay