*** MIDAS logfile (05SEPpl1.2) *** Su, 14 May 2006 16:37:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 May 2006 16:38:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060514 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 325.83 stddev = 4.42 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060514 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1987.7 1987.6 1629.6 -0.20 -0.21 1.22 2 2.13 3840.5 3839.9 3119.7 0.25 0.24 1.23 3 4.25 7511.7 7512.6 6124.7 0.07 0.08 1.23 4 6.38 11199.1 11199.0 8977.2 0.13 0.14 1.25 5 9.57 16714.9 16716.9 13434.3 0.10 0.11 1.24 6 12.76 22250.5 22250.3 17645.9 0.17 0.17 1.26 7 15.95 27756.2 27756.4 21638.4 0.11 0.11 1.28 8 21.27 36915.0 36915.0 28718.3 -0.01 -0.01 1.29 9 26.59 46023.3 46023.4 34993.9 -0.19 -0.19 1.32 10 31.91 55058.5 55056.5 41495.6 -0.44 -0.44 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2641 to 73154 e- Average count rate measured: 1728 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 5.88 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.44 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay