*** MIDAS logfile (05SEPpl1.2) *** Mo, 08 May 2006 22:44:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 08 May 2006 22:44:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060508 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 326.37 stddev = 5.59 FIERA temperature = 289.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060508 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1985.8 1985.0 1625.2 -0.19 -0.21 1.22 2 2.13 3837.8 3838.1 3112.5 0.24 0.25 1.23 3 4.25 7508.8 7509.3 5995.8 0.06 0.07 1.25 4 6.38 11196.1 11196.2 9697.8 0.13 0.14 1.15 5 9.57 16713.1 16712.1 13372.8 0.11 0.11 1.25 6 12.76 22247.4 22246.6 17737.1 0.17 0.17 1.25 7 15.96 27750.6 27752.8 21648.7 0.10 0.11 1.28 8 21.28 36909.7 36907.4 28445.4 -0.01 -0.01 1.30 9 26.59 46019.6 46017.8 35186.1 -0.18 -0.18 1.31 10 31.91 55049.7 55047.4 41120.2 -0.44 -0.44 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2658 to 73699 e- Average count rate measured: 1728 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.02 e-/ADU RON = 7.49 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.090 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.44 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay