*** MIDAS logfile (05SEPpl1.2) *** Sa, 06 May 2006 16:45:06 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 06 May 2006 16:46:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060506 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 324.65 stddev = 5.17 FIERA temperature = 289.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060506 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1991.1 1991.6 1747.5 -0.19 -0.21 1.14 2 2.13 3846.6 3846.1 3143.6 0.28 0.24 1.22 3 4.25 7520.6 7519.7 6172.9 0.07 0.07 1.22 4 6.38 11211.4 11210.3 8922.0 0.13 0.12 1.26 5 9.57 16732.8 16735.8 13251.4 0.10 0.12 1.26 6 12.76 22273.4 22275.0 17515.2 0.16 0.17 1.27 7 15.95 27786.4 27785.9 21742.2 0.10 0.11 1.28 8 21.27 36956.8 36953.5 29745.4 -0.01 -0.01 1.24 9 26.59 46076.2 46074.4 34850.9 -0.19 -0.18 1.32 10 31.91 55118.0 55117.4 41327.3 -0.45 -0.43 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2644 to 73172 e- Average count rate measured: 1730 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.02 e-/ADU RON = 6.86 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.44 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay