*** MIDAS logfile (05SEPpl1.2) *** Fr, 05 May 2006 16:05:44 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 05 May 2006 16:06:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060505 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 325.62 stddev = 6.92 FIERA temperature = 289.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060505 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1988.6 1988.8 1632.6 -0.15 -0.17 1.22 2 2.12 3842.5 3842.2 3156.9 0.27 0.27 1.22 3 4.25 7514.4 7514.3 6111.4 0.07 0.07 1.23 4 6.38 11201.0 11201.6 8944.3 0.13 0.13 1.25 5 9.57 16719.1 16720.1 13554.3 0.10 0.11 1.23 6 12.76 22255.7 22254.2 17655.2 0.16 0.15 1.26 7 15.95 27762.0 27762.9 21792.8 0.09 0.10 1.27 8 21.27 36923.9 36924.7 28867.8 -0.02 -0.02 1.28 9 26.59 46035.3 46034.9 34798.0 -0.20 -0.20 1.32 10 31.91 55073.1 55073.9 40953.3 -0.45 -0.45 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2667 to 73850 e- Average count rate measured: 1729 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.02 e-/ADU RON = 9.28 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.45 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay