*** MIDAS logfile (05SEPpl1.2) *** Fr, 21 Apr 2006 17:26:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 21 Apr 2006 17:27:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060421 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 324.86 stddev = 5.23 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060421 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1983.7 1984.2 1618.1 -0.17 -0.16 1.23 2 2.13 3838.6 3838.5 3107.6 0.28 0.27 1.24 3 4.25 7512.6 7512.1 6141.2 0.07 0.07 1.22 4 6.38 11201.4 11202.1 8997.3 0.13 0.13 1.24 5 9.57 16722.2 16723.5 13360.1 0.10 0.11 1.25 6 12.76 22260.5 22259.2 17579.5 0.16 0.15 1.27 7 15.96 27770.9 27770.6 21869.5 0.10 0.10 1.27 8 21.28 36935.2 36935.3 29003.5 -0.02 -0.02 1.27 9 26.59 46049.5 46050.4 35035.8 -0.20 -0.20 1.31 10 31.91 55092.1 55094.1 41122.5 -0.45 -0.44 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2644 to 73430 e- Average count rate measured: 1729 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.02 e-/ADU RON = 6.97 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.088 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.45 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay