*** MIDAS logfile (05SEPpl1.2) *** Sa, 15 Apr 2006 16:25:36 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 15 Apr 2006 16:26:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060415 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.93 stddev = 5.12 FIERA temperature = 289.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060415 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1998.5 1998.7 1640.5 -0.19 -0.21 1.22 2 2.12 3852.5 3852.4 3158.2 0.25 0.24 1.22 3 4.25 7528.1 7526.7 6054.6 0.07 0.07 1.24 4 6.38 11215.5 11216.2 8997.4 0.12 0.13 1.25 5 9.57 16737.4 16737.7 13496.6 0.11 0.11 1.24 6 12.76 22276.2 22273.9 17610.9 0.17 0.17 1.26 7 15.95 27786.3 27787.1 21602.5 0.11 0.11 1.29 8 21.27 36949.8 36950.0 28562.5 -0.02 -0.01 1.29 9 26.59 46068.6 46069.4 35069.0 -0.18 -0.18 1.31 10 31.91 55108.7 55108.8 40913.1 -0.44 -0.43 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2681 to 73927 e- Average count rate measured: 1729 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 6.87 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.098 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.44 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay