*** MIDAS logfile (05SEPpl1.2) *** Th, 06 Apr 2006 17:21:11 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 06 Apr 2006 17:22:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060406 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 327.07 stddev = 5.18 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060406 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1995.2 1995.1 1628.5 -0.23 -0.18 1.23 2 2.13 3855.2 3853.0 3158.9 0.29 0.27 1.22 3 4.25 7531.7 7533.2 6063.6 0.06 0.07 1.24 4 6.38 11226.5 11227.0 8919.8 0.13 0.13 1.26 5 9.57 16756.3 16754.7 13341.2 0.11 0.10 1.26 6 12.76 22300.4 22299.6 18400.2 0.17 0.16 1.21 7 15.95 27816.8 27817.2 22091.8 0.11 0.10 1.26 8 21.27 36994.3 36994.4 28729.6 -0.01 -0.02 1.29 9 26.59 46120.1 46123.0 35045.5 -0.18 -0.19 1.32 10 31.91 55173.6 55177.5 41198.3 -0.44 -0.44 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2662 to 73625 e- Average count rate measured: 1731 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.02 e-/ADU RON = 6.92 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.094 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.44 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay