*** MIDAS logfile (05SEPpl1.2) *** Mo, 20 Feb 2006 19:16:10 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 20 Feb 2006 19:17:08 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.16 stddev = 5.38 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2001.7 2001.1 1648.1 -0.16 -0.16 1.21 2 2.13 3872.8 3872.4 3133.6 0.27 0.26 1.24 3 4.25 7580.7 7580.7 6101.7 0.07 0.08 1.24 4 6.38 11303.0 11301.9 9029.3 0.13 0.13 1.25 5 9.57 16874.7 16873.6 13401.2 0.11 0.11 1.26 6 12.76 22461.6 22460.2 17670.6 0.16 0.16 1.27 7 15.96 28022.2 28020.3 21779.7 0.10 0.10 1.29 8 21.28 37267.5 37264.6 28852.3 -0.02 -0.02 1.29 9 26.59 46462.1 46460.5 35410.6 -0.20 -0.20 1.31 10 31.91 55583.8 55582.2 41666.6 -0.45 -0.45 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2665 to 74007 e- Average count rate measured: 1745 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 7.16 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.088 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.45 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay