*** MIDAS logfile (05SEPpl1.2) *** Th, 19 Jan 2006 22:45:03 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 19 Jan 2006 22:45:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060119 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.63 stddev = 6.25 FIERA temperature = 291.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060119 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 946.6 946.1 799.9 -0.05 0.02 1.18 2 2.00 1841.5 1841.2 1517.4 -0.00 0.01 1.21 3 4.00 3630.7 3630.4 2932.8 0.00 0.00 1.24 4 6.00 5420.0 5419.3 4408.9 0.01 -0.00 1.23 5 9.00 8103.8 8103.9 6450.1 0.02 0.02 1.26 6 12.00 10787.2 10786.8 9039.8 0.02 0.01 1.19 7 15.00 13470.7 13470.4 10777.7 0.02 0.01 1.25 8 20.00 17941.5 17942.1 14219.1 0.01 0.01 1.26 9 25.00 22406.8 22407.2 18337.9 -0.01 -0.02 1.22 10 30.00 26871.0 26870.9 21188.2 -0.04 -0.05 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1188 to 33745 e- Average count rate measured: 894 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.01 e-/ADU RON = 7.85 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.061 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay