*** MIDAS logfile (05SEPpl1.2) *** Mo, 09 Jan 2006 12:34:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 09 Jan 2006 12:35:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 060109 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 340.67 stddev = 3.82 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_060109 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1029.4 1028.8 845.0 -0.03 0.02 1.22 2 2.00 1968.8 1968.2 1591.7 -0.02 0.01 1.24 3 4.00 3846.1 3846.1 3148.6 -0.01 -0.01 1.22 4 6.00 5725.5 5725.9 4607.5 0.03 0.03 1.24 5 9.00 8542.4 8542.3 7952.1 0.03 0.02 1.07 6 12.00 11359.1 11359.9 9095.4 0.03 0.02 1.25 7 15.00 14174.5 14176.1 11242.8 0.02 0.01 1.26 8 20.00 18865.5 18868.6 14953.8 0.00 -0.00 1.26 9 25.00 23555.7 23555.6 18613.9 -0.01 -0.03 1.27 10 30.00 28237.6 28239.2 22029.2 -0.05 -0.07 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1324 to 36332 e- Average count rate measured: 939 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.02 e-/ADU RON = 4.91 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.099 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay