*** MIDAS logfile (03SEPpl1.2) *** We, 21 Sep 2005 16:32:01 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 21 Sep 2005 16:32:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050921 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 358.50 stddev = 3.92 FIERA temperature = 290.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050921 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1045.9 1046.4 864.1 -0.04 -0.02 1.21 2 2.00 2006.2 2005.1 1625.3 0.00 -0.01 1.23 3 4.00 3921.5 3922.1 3453.8 -0.00 0.00 1.14 4 6.00 5840.2 5839.9 4701.5 0.02 0.01 1.24 5 9.00 8715.9 8716.2 7187.8 0.03 0.03 1.21 6 12.00 11591.8 11591.3 9300.8 0.03 0.02 1.25 7 15.00 14466.3 14466.7 11536.3 0.02 0.02 1.25 8 20.00 19257.5 19258.0 15284.0 0.01 0.01 1.26 9 25.00 24043.3 24043.2 19452.4 -0.01 -0.02 1.24 10 30.00 28823.4 28825.3 22802.3 -0.05 -0.05 1.26 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1321 to 36398 e- Average count rate measured: 958 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.26 +/- 0.01 e-/ADU RON = 4.95 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.095 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay