*** MIDAS logfile (03SEPpl1.2) *** We, 07 Sep 2005 15:34:25 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 07 Sep 2005 15:35:27 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050907 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 359.78 stddev = 4.91 FIERA temperature = 290.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050907 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1045.6 1045.8 869.8 -0.04 -0.03 1.20 2 2.00 2000.5 2000.7 1657.2 -0.03 -0.02 1.21 3 4.00 3912.1 3911.9 3134.2 0.02 0.01 1.25 4 6.00 5822.7 5823.0 4698.2 0.02 0.03 1.24 5 9.00 8688.2 8688.2 6985.4 0.03 0.02 1.24 6 12.00 11552.8 11553.2 9201.3 0.02 0.02 1.26 7 15.00 14418.1 14418.8 11584.1 0.02 0.02 1.24 8 20.00 19192.0 19192.3 15196.2 0.01 0.01 1.26 9 25.00 23960.1 23960.6 18934.2 -0.01 -0.01 1.27 10 30.00 28722.8 28723.3 23392.9 -0.05 -0.05 1.23 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1304 to 35804 e- Average count rate measured: 955 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.01 e-/ADU RON = 6.12 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.098 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.03/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay