*** MIDAS logfile (03SEPpl1.2) *** We, 06 Jul 2005 19:04:07 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 06 Jul 2005 19:05:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050706 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 361.72 stddev = 7.33 FIERA temperature = 290.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050706 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1052.4 1052.4 913.3 -0.03 0.01 1.15 2 2.00 2017.1 2017.5 1688.6 -0.03 0.01 1.19 3 4.00 3948.0 3947.7 3906.9 0.02 0.02 1.01 4 6.00 5878.4 5878.0 4764.2 0.02 0.01 1.23 5 9.00 8773.1 8773.7 7076.6 0.02 0.02 1.24 6 12.00 11668.4 11669.1 9414.3 0.03 0.02 1.24 7 15.00 14562.5 14563.0 11594.5 0.02 0.01 1.26 8 20.00 19386.1 19387.0 15488.5 0.01 -0.00 1.25 9 25.00 24205.0 24204.6 19079.8 -0.01 -0.03 1.27 10 30.00 29017.3 29017.1 22699.5 -0.05 -0.07 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1358 to 37434 e- Average count rate measured: 965 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 9.46 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay