*** MIDAS logfile (03SEPpl1.2) *** Th, 23 Jun 2005 17:41:37 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 23 Jun 2005 17:42:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050623 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 360.42 stddev = 6.88 FIERA temperature = 290.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050623 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1071.7 1071.6 921.0 0.01 -0.01 1.16 2 2.00 2046.7 2046.9 1696.8 -0.02 -0.01 1.21 3 4.00 3998.3 3997.7 3311.5 0.02 0.01 1.21 4 6.00 5949.6 5949.9 4814.8 0.02 0.02 1.24 5 9.00 8875.3 8875.9 8185.0 0.01 0.02 1.08 6 12.00 11802.7 11802.0 9549.7 0.02 0.02 1.24 7 15.00 14727.2 14727.5 11660.8 0.01 0.02 1.26 8 20.00 19603.5 19602.2 15605.2 0.01 0.01 1.26 9 25.00 24472.6 24473.0 19292.8 -0.02 -0.02 1.27 10 30.00 29334.2 29334.3 23177.1 -0.06 -0.06 1.27 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1371 to 37528 e- Average count rate measured: 975 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.02 e-/ADU RON = 8.80 +/- 0.14 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.101 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay