*** MIDAS logfile (03SEPpl1.2) *** Th, 28 Apr 2005 15:56:00 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Th, 28 Apr 2005 15:56:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050428 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 358.52 stddev = 6.87 FIERA temperature = 290.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050428 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 985.6 985.3 843.6 0.02 -0.04 1.17 2 2.00 1914.9 1914.4 1572.6 0.01 -0.01 1.22 3 4.00 3772.5 3773.1 3084.7 -0.00 -0.00 1.22 4 6.00 5631.0 5630.8 4576.7 0.01 0.02 1.23 5 9.00 8419.4 8418.6 6734.3 0.02 0.02 1.25 6 12.00 11206.5 11205.3 8942.0 0.02 0.02 1.25 7 15.00 13993.9 13993.2 11289.9 0.01 0.03 1.24 8 20.00 18635.4 18635.8 14816.8 -0.00 0.02 1.26 9 25.00 23273.7 23273.1 18315.5 -0.03 -0.01 1.27 10 30.00 27907.9 27909.0 22127.7 -0.06 -0.04 1.26 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1250 to 35410 e- Average count rate measured: 929 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.00 e-/ADU RON = 8.71 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.063 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay