*** MIDAS logfile (03SEPpl1.2) *** Fr, 04 Mar 2005 16:20:50 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 04 Mar 2005 16:22:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 050304 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.85 stddev = 6.40 FIERA temperature = 291.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rA2_050304 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1072.5 1071.8 903.8 -0.02 -0.05 1.19 2 2.00 2058.3 2058.3 1691.7 0.00 0.01 1.22 3 4.00 4029.4 4029.3 3249.5 0.00 0.00 1.24 4 6.00 6001.4 6001.0 4972.6 0.02 0.02 1.21 5 9.00 8958.1 8958.1 7203.6 0.02 0.02 1.24 6 12.00 11915.1 11915.2 10221.5 0.02 0.02 1.17 7 15.00 14871.6 14871.8 11876.7 0.02 0.02 1.25 8 20.00 19797.0 19796.9 15665.9 0.01 0.01 1.26 9 25.00 24719.3 24720.0 19395.1 -0.01 -0.01 1.27 10 30.00 29629.3 29630.5 23026.3 -0.06 -0.06 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1381 to 38163 e- Average count rate measured: 985 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.02 e-/ADU RON = 8.24 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.091 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.02/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay